MI/B&C Observing Procedures
1. Accessing Data Acquisition Computer (LLama)
- Username: obsmag1b
- Password: check with the staff
- Open an xterm window
- cd /home/obsmag1b/data
- mkdir 20020914_15 (e.g.)
- cd 20020914_15
2. Start Detector Control Window in LLama
- From xterm window type 'ccd' to open detector control window
- Before the CCD control window appears you have to select 4 parameters from the menu:
- Uniblitz
- Marconi1
- 50
- magellan1
- On the CCD control window select Gain 1
- On the CCD control window select Subraster and enter the following values:
- 1 2048 140 480
- Save spectra.subr
- Apply
- Done
- Open an IRAF window and go to the recently created directory (20020914_15)
- Load the 'bnc' package twice
- Open XImtool
- In IRAF window reset stdimage=imt27
3. Open Flat Field Screen Lamp Window
- From any window type 'ffs' to open Flat Field Screen Lamp Window
- From any window type 'bnc' to open bnc GUI
4. Check instrument setup
- Set Decker to 35, and Prism to 0 in bmc GUI
- Take a He Ar lamp to make sure that wavelength range is correct. Adjust
grating tilt accordingly
- Check that arc lines are properly aligned with columns
- Check focus of spectrograph (near collimator value of 925)
5. Dome Flats
- Set Decker to 35, and Prism to 100 in bmc GUI, and micrometer to 300 microns
- Turn ON quartz light at the platform and adjust voltage
- Turn OFF platform lights
- In Flat Field Screen Lamp Window click IN to move screen into place
- In CCD window set exposure time and object to "Dome Flat", imtype="flat"
- Exposure times:
- 300u slit : 60 sec (voltage=4.0)
- 500u slit : 25 sec (voltage=4.0)
- Start sequence with START botton
- When the sequence of flats is complete turn OFF quartz lamp
- In Flat Field Screen Lamp Window click OUT to move screen into place
- Take a sequence of bias frames (Set Exptime=0, Imtype=zero, Loop=15)
6. Observing
- Set Imtype=object
- ASk NA to enable PA automatic rotation
- In CCD window set exposure time, object to "object name", and click START to begin exposure
- To take arc lamp click IN to move screen into place, click He and Ar (both of them), and take a 20-sec arc exposure (don't forget to remind the NA to disable guider)
- Click OUT to move screen into place, and click He and Ar to turn OFF the lamps
7. Before going to bed
- After finishing the darks turn OFF monitors in control room
- Complete Observing Run Report Form
mhamuy@lco.cl